![Circassia NIOX VERO User Manual Download Page 47](http://html.mh-extra.com/html/circassia/niox-vero/niox-vero_user-manual_2607894047.webp)
Chapter 11 Technical data
000190-09 NIOX VERO
®
User Manual US
45
11.14.3 FCC
FCC CAUTION
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference
that may cause undesired operation.
Properly shielded and grounded cables and connectors must be used for con-
nection to host computers and / or peripherals in order to meet FCC emission
limits.
Note: This equipment has been tested and found to comply with the limits for
a Class B digital device, pursuant to part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful interference in
a residential installation. This equipment generates, uses and can radiate ra-
dio frequency energy and, if not installed and used in accordance with the in-
structions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or tele-
vision reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more of
the following measures:
•
Reorient or relocate the receiving antenna.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
•
Consult the dealer or an experienced radio/TV technician for help.
Contains FCC ID: SQK-MBH7BTZXXC2
11.15
Rechargeable battery capacity
Only use the power adapter or USB cable supplied by Circassia to charge the
battery.
Capacity: Approx. 30 measurements per day or 36 hours stand-by in 77°F en-
vironment condition.
Lifetime: At least one year with normal use.
Charging time: <8 hours under normal conditions.
Lowered capacity, and/or when 8 hours of charging time does not charge the
battery fully, indicates that the battery should be replaced.
Battery Type No BJ-G510039AA, Article No 12-1150.
Note:
To charge the battery by USB cable, the instrument needs to be
powered off
.
11.16
Instructions for transport and storage
CAUTION!
Always use a closed bag or box for transportation and storage of
NIOX VERO
®
.
1.
Verify that the instrument is turned off and disconnected from the power
supply.
2.
Remove the patient filter and attach the protective cap on the handle.
3.
Place the instrument and accessories in the bag and close bag.
4.
Verify that the storage environment conditions are appropriate (see
recommendations for NIOX VERO, including Sensor.
Summary of Contents for NIOX VERO
Page 54: ......