Chapter 9 Safety information
40
000190-09 NIOX VERO
®
User Manual US
9.5
Emission of electromagnetic energy
CAUTION!
This equipment has been designed and tested to CISPR 11
Class B. In a domestic environment it may cause radio interference, in which
case, you may need to take measures to mitigate the interference.
For more information see “Electromagnetic emissions” on page 51.
9.6
Operating conditions
Ensure stable operating conditions by avoiding placement of the instrument
in direct sunlight, near sources radiating heat, or ventilation. NIOX VERO op-
erates during the following conditions:
•
NO in ambient air up to 300 ppb
To verify NO in ambient air, perform an ambient measurement, see page 15
•
Exhaled flow during FeNO measurements at 50 ml/s (3 L/min) ± 10%
BTPS* during 10 seconds.
*Body Temperature Pressure Saturated, 98.6 °F (37 °C), 14.7 psi
(1013 hPa), 100% RH
•
Temperature range of 50°F to 95°F
•
A relative humidity range of 20% to 80%, non condensing
•
An atmospheric pressure range of 700 hPa to 1060 hPa
Performance shall be sustained when measuring continuously at a rate of
up to 10 measurements / hour.
Exhaled breath contains water vapor which can condense inside the in-
strument. When excessively used in a short period, there is a risk for
condensation of water inside NIOX VERO.
Normally a maximum of 10 exhalations/hour can be performed with
NIOX VERO during continuous use. However, it is possible to perform 20 ex-
halations in one hour if the instrument is paused for a minimum of 30 minutes
prior to the next session of exhalations. Exhalations include failed and suc-
cessful measurements.
9.6.1
Limited warranty
Circassia provides a Limited Warranty for this instrument and original acces-
sories delivered with the instrument. Conditions are defined when the items
are purchased.
Do NOT try to repair the instrument. Any attempt will make the warranty in-
valid and performance according to the specifications cannot be guaranteed.
9.6.2
Support
Contact your local Circassia representative or Circassia Technical Support if
you encounter problems which you cannot solve with the information in this
manual.
For contact details, see back cover, and provide the following information:
•
Your name, address and telephone number.
•
Serial number for both the instrument, handle and Sensor.
•
Problem description (as thorough as possible)
•
Alert codes or lists.