Chapter 5 Using NIOX VERO
®
000190-09 NIOX VERO
®
User Manual US
21
Note:
During the qualification days of a new QC tester the result is
displayed as presented below.
(A) Date of measurement, (B) Status button (green = within specifications,
white = not within specifications), (C) Mean value in ppb, (D) QC tester
number
Press
to return to main menu.
Repeat the QC test if the positive and/or the negative control fail. If the QC
failure persists, discontinue use of NIOX VERO
®
and contact
Circassia Technical Support.
Note:
The prompt to QC the device will remain if the QC measurement was
performed by a non-qualified QC candidate. It is not an indication of
an unsuccessful (failed) QC measurement.
Note:
If the daily Quality Control is not
successfully performed, or if the
results from the QC are outside
limits, a warning alert will be
displayed beside the measurement
value.
5.8.3
View QC logs
1.
Select the
Settings
button on the main menu.
2.
Select the
QC log
button.
The QC log window opens.
3.
Browse through the measurement logs using the backward and forward
buttons.
5.8.4
View QC information
After day 3 the QC qualification procedure is complete. A qualified tester is
displayed in the QC tester view