Chapter 7 Troubleshooting
28
000190-09 NIOX VERO
®
User Manual US
7
Troubleshooting
7.1
Alert codes and actions
Alert messages and other information are shown as a code on the instru-
ment display. The tables below provide the alert codes and recommended
actions to be taken for an alert code. If the alert persists, contact your local
Circassia representative or Circassia Technical Support.
User alerts
Screen
Action
A10
Exhalation too strong
Press Return and repeat the
measurement with less exhalation force.
A11
Exhalation too weak
Press Return and repeat the
measurement with greater exhalation
force.
A12
Measurement failed
No exhalation detected or the user failed
to exhale within 15 seconds from inhaling.
Press Return button, restart the
measurement and exhale into the
instrument directly after inhalation.
A13
Analysis interrupted
Repeat the measurement and do not
breathe through the handle during
analysis.
A21
Measurement failed
Remove any sources of disturbance
(such as cordless phones/mobile phones
or gas emitting appliances). Then press
Return. When the instrument is ready for
use repeat the measurement.
If the alert persists, unplug and reconnect
the power supply unit to restart the
instrument.
Instrument
alerts
Screen
Action
A01
Unstable temperature
Make sure that the ambient temperature
is between 50°F and 95°F. Wait for the
Sensor to stabilize. If necessary move
the instrument to another location and
restart the instrument.
A02
Sensor stabilization
Remove any sources of disturbances
(such as cordless phones, mobile
phones or gas emitting appliances).
Wait for the Sensor to stabilize.