Chapter 7 Troubleshooting
32
000190-09 NIOX VERO
®
User Manual US
A56
Handle cap time out
Failure to press the forward button in
time (within 2.5 min.).
Repeat the QC measurement and make
sure to press the forward button
immediately after the Handle Cap is
attached.
A57
NO scrubber analysis error
NO scrubber analysis phase error.
Restart the QC measurement.
A80
The instrument is about to expire
Order a new instrument. This alert is
visible when less than 500
measurements remain or less than 120
days until expiry date.
Press OK to acknowledge.
A81
The Sensor is about to expire
Order a new Sensor. This alert is visible
when less than 10% of the
measurements remain or less than 2
weeks until expiry date.
Press OK to acknowledge.
Instrument
alerts
Screen
Action
A82
The breathing handle is about to
expire
This alert is visible when less than 100
measurements remain or less than 2
weeks until expiry date.
Press OK to acknowledge. Prepare to
change breathing handle.
A90
All measurements on the instrument
have been used
It is still possible to view measurements
stored in the instrument memory.
A91
All measurements on the Sensor
have been used.
Replace the Sensor, see page 35.
A92
Instrument expiration date has
passed
It is still possible to view measurements
stored in the instrument memory.
Instrument
alerts
Screen
Action