![Circassia NIOX VERO User Manual Download Page 46](http://html.mh-extra.com/html/circassia/niox-vero/niox-vero_user-manual_2607894046.webp)
Chapter 11 Technical data
44
000190-09 NIOX VERO
®
User Manual US
11.10
Exhalation parameters
Exhalation time: Standard mode: 10 s (clinical use)
All exhalations are to be performed at an exhalation pressure of 10 - 20 cm
H
2
O, to maintain a fixed flow rate of 50 ±5 ml/s. The instrument stops the
measurement at pressures outside the interval. Warning alerts sounds at
10 - 12 and 18-20 cm H
2
O.
11.11
Essential performance
Essential performance for NIOX VERO consists of
1.
The measurement of the NO concentration of exhaled human breath
2.
The control of exhaled breath for Asthma management according to
ATS/ERS
NIOX VERO contains internal monitoring functionality for safety and
essential performance parameters. Recurrent testing is not necessary to
maintain essential performance or basic safety.
11.12
Memory capacity
Up to 15, 000 measurements, depending on the size of the measurement
files.
11.13
Patient filter
Disposable bacterial/viral filter to be changed for each patient.
11.14
Bluetooth
NIOX VERO
®
has a Bluetooth class 2 receiver/transmitter with:
•
Frequency band of 2402MHz~2480 MHz.
•
Modulation method
• 0,5BT Gaussian FIlter 2 FSK modulation index: 0.28~0.35 (Basic
Rate 1Mbps)
•
π
/4-DQPSK (EDR 2Mbps)
•
8DPSK (EDR 3Mbps)
•
ERP
• Power class 2
11.14.1 R&TTE Directive
Hereby, Circassia AB, declares that this NIOX VERO is in compliance with
the essential requirements and other relevant provisions of Directive 1999/5/
EC.
11.14.2 IC
This Class B digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe B est conforme à la norme NMB-003 du
Canada.
This device complies with Industry Canada license-exempt RSS stan-
dard(s). Operation is subject to the following two conditions: (1) this device
may not cause interference, and (2) this device must accept any interfer-
ence, including interference that may cause undesired operation of the de-
vice.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables
aux appareils radio exempts de licence.
L'exploitation est autorisée aux deux conditions suivantes: (1) l'appareil ne
doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter
tout brouillage radioélectrique subi, même si le brouillage est susceptible
d'en compromettre le fonctionnement.
Contains IC: 337L-MBH7BTZXXC2
Summary of Contents for NIOX VERO