70
SmartMonitor 2 PS Parents’ Guide
P
roblem
P
ossible
c
ause
i
nstruc tions
Alarm Sound Continuous, No Lights. Incorrect power-off sequence.
Internal part failure.
•
Press the POWER button, and ensure that the power light
is illuminated.
• P
ress and hold the RESET button. Press and release the
POWER button. Wait two seconds. Then release the RESET
button.
Contact your home care provider.
Alarm Sound Continuous, No Lights. No power, battery drained.
Connect power supply. Use Power-Off to silence alarm.
•
Press and hold the blue RESET button.
•
Press and release the gray POWER button. Wait five
seconds. Then release the RESET button.
Prior to use, allow the battery to charge approximately
30 minutes. You may then operate the monitor while it is
plugged in. Allow the battery to charge for 6 hours before
using the monitor on battery power.
Contact your home care provider.
Alarm sounds weak.
Internal part failure.
Low battery.
Contact your home care provider.
Charge battery.