SmartMonitor 2 PS Parents’ Guide
i
mmunit y
t
est
iec 60601 t
est
l
evel
c
omPliance
l
evel
e
lec tromagnetic
e
nvironment
-
g
uidance
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV for common mode
Mains power quality should be that of a typical home or hospital environment.
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
<5% U
T
(>95% dip in U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
Mains power quality should be that of a typical home or hospital environment.
If the user of the device requires continued operation during power mains
interruptions, it is recommended that the device be powered from an
uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical
location in a typical hospital or home environment.
NOTE:
U
T
is the AC mains voltage prior to application of the test level.