SmartMonitor 2 PS Parents’ Guide
27
r
esPiration
, h
eart
,
and
s
P
o
2
m
onitoring
After you unpack your monitor and make sure you have all the parts, follow the steps listed below to set it up.
Step 1: Set the Monitor on a Clean, Flat Surface.
•
Be sure the speaker is not blocked.
•
To avoid interference, be sure that no other electrical appliances are within three feet of the unit.
•
Make sure the monitor is close enough to connect to the child comfortably.
WARNING:
Do not place the monitor in bed with a child.
Step 2: Connect the ECG Patient Cable to the Monitor.
See the illustration at the end of this step.
•
Insert the round end of the ECG patient cable into the bottom round connector found on the front of the
monitor.
•
Insert the connector with the red dot facing up. The connector will snap into place.
•
To remove the ECG patient cable, grasp it at the end of the patient input connector and gently pull back.
You should feel the connector body slide back and unlock the connector as you pull.
CAUTION:
Do not twist or turn the ECG patient cable to remove it from the monitor as this may damage the ECG patient
cable and/or monitor.
CAUTION:
Do not place the ECG patient cable over the top of the crib rail. The cable should be placed between the vertical
bars.