5. Set the required temperature via the SR-800R Controller and wait until it reaches
a stable temperature (a message is displayed:
Waiting for Refrigerator
).
Note that the Circulator temperature is set automatically by the SR-800R
Controller and according to
Table 8-1
:
Table 8-1:
SR-800R-to-Refrigerator Temperature Range
Refrigerator SR-800R
Temp. Range
-30ºC
-40ºC to -20ºC
-10ºC
-20ºC to 0ºC
+20ºC
0ºC to +40ºC
+60ºC
+40ºC to +80ºC
+100ºC
+80ºC to +150ºC