CI Systems
SR-800R
7. Removable
Sensors
This chapter describes the Removable Emitter Sensor removal and installation
procedures that are performed on two types of blackbodies.
IMPORTANT
It is essential to update the calibration tables after replacing the
Emitter and/or Temperature sensors!
7.1 General
The process of the Removable Sensor replacement includes two main stages:
Replacement of the Sensor
Update of the Calibration Table (a must!)
The following sections provide the procedures:
Section
7.2: Emitter Sensor Replacement
Section
7.3
:
Target Sensor Replacement
Section
7.4
:
Updating the Calibration Tables
CAUTION
Shut down the power supply prior to any action taken. Allow
adequate time for the Emitter to cool down.
Avoid bending the sensor wire and the connections
between wire to sensor/USB connector. The sensor is fragile -
take care.
605-7040
Operation
Manual
Rev D
7-1
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for ControlMaster SR-800R
Page 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 129: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...