CI Systems
SR-800R
7.3.2.2 Installation
1. Clean the Sensor port using the plastic straw (supplied with the new sensor):
a. Insert the plastic straw into the Sensor port.
b. Rotate the plastic straw several times in the Sensor port.
c. Remove the plastic straw from the Sensor port.
2. Apply a thin layer of thermal grease (supplied with the new sensor) on the bottom
section of the sensor (approx. 1cm on the free end of the sensor), using a
toothpick (see
Figure 7-6
).
3. Insert the sensor into the port until the magnet secures it (see
Figure 7-10
).
4. Connect the Sensor cable-connector (see
Figure 7-7
).
5. Arrange the sensor cable.
605-7040
Operation
Manual
Rev D
7-13
June 2013