Q9 Digital Color Doppler Ultrasound System
23
WARNING:
To avoid risk of fire, the system power must be supplied from a separate, properly rated
outlet.
Under no circumstances should the AC power plug be altered, changed, or adapted to a
configuration rated less than specified. Never use an extension cord or adapter plug.
To help assure grounding reliability, connect to a “hospital grade” or “hospital only”
grounded power outlet.
3.4.3.
Power On
NOTE:
Turn on the green power switch (main power circuit breaker switch, see Fig. 3-1 c in Section
3.2.1 Console Overview) at the back of the system, and then press the Power button on the left
of control panel to turn on the system.
Power up Sequence:
The system is initialized and start-up status is reflected on the monitor:
control panel flashing and then getting dark
system checking BIOS data
booting the operation system
loading software
entering examination status
HINTS
The power up procedure takes about approx. 180 seconds. If a problem occurs, take a picture
and record the error information for service reference.
NOTE:
While the system is on, DO NOT fold the keyboard.
While unfolding the keyboard, please hold and place the keyboard slowly and lightly on
the desk.
After power off the system, please wait for more than 3 minutes to power on again.
When the system is powered on, for safety reason, please avoid the following
-
close the keyboard
-
move the system
3.4.4.
Power Off
To power off the system:
Press the Power button on the left of control panel.
When the screen shows “Turn Off”, “Restart”, and “Cancel”, press “Turn off” to shutdown the system.