Q9 Digital Color Doppler Ultrasound System
165
4. After click “Chison System Recovery”, the system will start to backup ultrasound system data, after finishing the
backup, the system will transfer to the recovery procedure automatically, and the following interface will pop up.
Click “OK” to confirm and start the recovery.
Click “Cancel” to cancel the system recovery, and go back to the CMD interface.
5. After finishing the recover, the system will pop the CMD interface. Input EXIT in the command line to reboot the
system will reboot.
Note:
After recovering the system successfully, users need to set "display Settings" manually to ensure video
printer work properly
Setting procedures are as follows:
Press “Set-up” key to enter into the setting interface, select the system interface, and click “display setting”