Q9 Digital Color Doppler Ultrasound System
157
Chapter 9 System Maintenance and Troubleshooting
9.1 Back up information
Caution:
All patient data created is NOT backed-up! It is highly recommended to create a full system backup of
patient data regularly and empty the hard disk (HDD), to ensure the hard disk (HDD) has never
reached its maximum capacity.
NOTE:
To Backup exams to USB DVD/CD+(R)W disk, please confirm that the USB DVD/CD+(R)W storage
medium used is clean and not scratched!
Warning:
Do not disconnect an external USB pen drive without stopping it. Disconnecting without stopping can
lead to data loss on the external device.
9.2 System Care and Maintenance
The system is a precise electrical device. To ensure the best performance and operation of the system, observe
proper maintenance procedures. Contact the local CHISON’s Authorized Service Representative for parts or
periodic maintenance inspections.
Inspecting the System
Examine the following on a monthly basis:
Connectors on cables for any mechanical defects.
Entire length of electrical and power cords for cuts or abrasions.
Equipment for loose or missing hardware.
Control panel and keyboard for defects.
To avoid electric shock hazard, do not remove panels or covers from console. This servicing must be performed by
CHISON's authorized service engineer. Failure to do so could serious injury.
If any defect is observed or malfunctions occur, do not operate the equipment but inform CHISON’s authorized