Q9 Digital Color Doppler Ultrasound System
7
VIDEO, S–VIDEO and VGA are not designed to work at the same time, only one port can work
at one time. Please select your most suitable port.
Acceptable Devices
The Patient Environmental devices shown above are specified to be suitable for use within the PATIENT
ENVIRONMENT.
CAUTION:
DO NOT connect any probes or accessories without approval by CHISON within the PATIENT
ENVIRONMENT.
DO NOT touch patient and devices without IEC/EN 60601-1 approval to avoid the leakage
current risk within the PATIENT ENVIRONMENT.
Unapproved Devices
CAUTION:
DO NOT use unapproved devices.
If devices are connected without the approval of CHISON, the warranty will be INVALID.
The system can’t be used with HF surgical equipment; otherwise the burns to patient may occur.
Any device connected to this system must conform to one or more of the requirements listed
below:
IEC standard or equivalent standards appropriate to devices.
The devices shall be connected to PROTECTIVE EARTH (GROUND).
CAUTION:
Unsafe operation or malfunction may occur. Use only the accessories, options and
supplies approved or recommended in these instructions for use.
Peripheral used in the patient environment
The system has been verified for overall safety, compatibility and compliance with the following on-board image
recording devices:
B/W video printer: Mitsubishi P93W; Sony UP-897MD
,
Sony UPD-711MD
Color video printer: Mitsubishi CP31W
The system may also be used safely while connected to devices other than those recommended above if the devices and
their specifications, installation, and interconnection with the system conform to the requirements of IEC/EN 60601-1-
1.
The connection of equipment or transmission networks other than as specified in the user instructions can result in an