Q9 Digital Color Doppler Ultrasound System
156
usable).
When cleaning the TV probe, it is important to be sure that all surfaces are thoroughly
cleaned.
Disinfecting
2% Glutaraldehyde-based solutions have been shown to be very effective for this purpose. Cidex is the only
germicide that has been evaluated for compatibility with the material used to construct the probes.
To keep the effectiveness of the disinfection solutions, a thoroughly cleaning must be done to the probe before
the disinfecting, make sure no residues remain on the probe.
Disinfecting Procedure:
Following all precautions for storage, use and disposal, prepare the germicide solution according to the
manufacturer’s instructions.
Place the cleaned and dried probe to contact with the germicide, being careful not to let the probe drop to the
bottom of the container and thus damage the probe.
After placing/immersing, rotate and shake the probe while it is below the surface of the germicide to eliminate
air pockets. Allow the germicide to remain in contact with the fully immersed probe. For high level
disinfection, follow the manufacturer’s recommended time.
Following all precautions for storage, use and disposal, prepare the germicide solution according to the
manufacturer’s instructions.
After removing from the germicide, rinse the probe according to the germicide manufacturer’s rinsing
instructions.
Flush all visible germicide residues from the probe and allow to air dry.