Q9 Digital Color Doppler Ultrasound System
17
3.1.2.
Transport and Storage Environmental Requirements
The following environmental transport and storage conditions are within system tolerances:
Temperature:
-5º C ~ 40º C
Relative Humidity:
≤
80% non-condensing
Atmosphere Pressure: 700hPa ~ 1060hPa
3.1.3.
Electrical Requirements
Power Requirements
AC 100-240V, 50-60Hz
Fuse Requirements
Fuse specification is 250V, 5.0 A (time-lag), the model is 50T T5AL 250V
CAUTION:
Please use the fuse provided by manufactory if the fuse need replace. User can’t buy and exchange the fuse
by themselves.
Power Consumption:
300 VA
Voltage Fluctuation
WARNING:
Maintain a fluctuation range of less than ±10% of voltage labeling on rear panel of the
system, otherwise the system may be damaged.
Grounding
Before connecting the power cord, connect the attached ground protection cable from Equipotentiality
terminal on system rear panel to a specialized grounding device.
NOTE:
Please follow the outlined power requirements. Only use power cords that meet the system
guidelines—failure to follow these procedures may result in system damage.
Line power may vary in different geographic locations. Refer to the detailed ratings on the
rear panel of the system for detailed information.