background image

INST.No.INE-454-P2 Software 

Ver.1.2

 

- 14 - 

 

4.7 Automatic emissivity calculation type selection 

(External setting :Model IR-CAQ

5

 only)

 

This mode is for the selection of the automatic emissivity calculation function (disable or enable). 
In case of using the automatic emissivity calculation functions, select the internal or external setting. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

4. Engineering mode 

1) To select the automatic emissivity calculation type, press 

SEL

 key in the engineering mode screen until “

E.Aut

” 

appears in the sub display

 

as shown in the right figure.   

2) By pressing 

 key, “

no

”, “

in

” or “

Et

” will blink in the 

main display.   

3) Use 

or 

key to select the desired type. 

4) To activate this selection, press 

ENT

 key.

 

    Tb   CONT   MEM   PEAK

 AL

AH

C
F

ε

[Calculation disable] 

Main display 

Sub display

    Tb   CONT   MEM   PEAK

 AL

AH

C
F

ε

[External automatic Emissivity 
 

 

calculation(by analog input)] 

Main display

Sub display

    Tb   CONT   MEM   PEAK

 AL

AH

C
F

ε

[Internal automatic Emissivity 
 

 

calculation(by keys)] 

Main display

Sub display

[Automatic emissivity   

calculation types] 

no

: Calculation disable 

in 

: Internal setting (by key) 

Et

: External setting (by analog input) 

 The default is 

no 

(automatic emissivity calculation disable). 

R e m a r k s

Summary of Contents for IR-CA series

Page 1: ...ware Ver 1 2 IR CAQ Series HIGH FUNCTION TYPE INFRARED RADIATION THERMOMETER Software volume Model IR CAQ Always keep these instruction with the unit Please be sure to deliver these instructions with the unit to the end user ...

Page 2: ...manual The symbols shown below are used depending on important degrees for using the thermometer safely and avoiding unexpected situations Important degree Symbols Contents 1 This symbol is attached to a title for the sentence with 2 For avoiding dangerous accidents may cause death or serious injury like as electrical shock fires or troubles damages of the thermometer 3 For avoiding injury or in p...

Page 3: ... 16 4 9 1 2 Zero side adjustments Adjustment by furnace temperature at zero side 16 4 9 1 3 Span side adjustment Adjustment by furnace temperature at span side 16 4 9 2 Zero span adjustments 2 17 4 9 2 1 Zero span adjustments execution For 2coL 17 4 9 2 2 Zero span adjustments execution For 1coL 17 4 9 3 Correction after the Zero span adjustments measured value became abnormal 17 4 9 3 1 Zero span...

Page 4: ...g 2 2 1 Parameter programming selection Operating mode In the measurement mode program or select a parameter by using SEL key The operator mode covers programming or selection of emissivity automatic emissivity calculation signal modulation mode modulation degree and alarms For each program menu refer to 3 Operating mode The right figure shows the flow chart of the parameter menus by pressing SEL ...

Page 5: ...mode Peakhold Samplehold Reset type Analogoutput Low temperature Analogoutput High temperature Inprogrammingof parameters thescreenwill movetothemeasurement modeunlessanykeyispushedfor 1minute Reset time Analogdummy output Contact output type Temperature unit Automatic emissivity calculation enable disable Int Ext Analogoutput correction Number of data Temperature before correction Temperature aft...

Page 6: ... the emissivity ratio can be set up separately at low temperature side range and high temperature side range Emissivity ratio programming 1 Emissivity ratio programming 2 Measurement wavelength Wavelength display goes off Wavelength display 1ligths 2col 2 color measurement automatic range switching 1 55 1 35µm emissivity ratio 1 35 0 9µm emissivity ratio 1col Single color measurement automatic ran...

Page 7: ...T key 3 1 2 Emissivity ratio programming 2 Program the emissivity ratio of 1 35 0 9µm in 2col and 0 9µm in 1col in case of automatic emissivity calculation In case of automatic emissivity calculation program the emissivity ratio at high temperature side range Even if it sets up in case of 2 color measurement and single color measurement it is not related to measurement 3 2 Automatic emissivity cal...

Page 8: ... by the original signal When the measured value is falling down the measured value is displayed by the damping degree programmed in 3 4 2 Damping degree selection Tb CONT MEM PEAK AL AH C F ε Sub display Main display Necessary to program the scaling of the analog input in 4 11 Analog input programming Reference 1 To select the external programming press SEL key in the measurement mode screen until...

Page 9: ...e or key to program the desired value 4 To shift to the next lower digit press key The next lower digit will blink 5 Repeat the above procedure of 3 and 4 up to the least significant digit for programming 4 digits 6 To activate this programming press ENT key AL AH C F ε Tb CONT MEM PEAK Sub display Main display By selecting dELy and programming to 0 0 the measured value based on the original signa...

Page 10: ...alarm disable and the programming an alarm set point 1 point only in the high alarm or the low alarm 3 5 1 Alarm type selection High alarm low alarm or alarm disable Alarm type No Alarm disable Hi High alarm The open collector of the contact output is ON when the measured value exceeds the alarm set point Lo Low temperature alarm The open collector of the contact output is OFF when the measured va...

Page 11: ...ing mode screen in 6 3 Parameter programming selection Engineering mode 4 1 Analog output scaling programming 4 1 1 Analog output minimum value programming OutL This programming is for the minimum value of the temperature scaling when the measured value is outputted by analog signal 4 1 2 Analog output maximum value programming OutH This programming is for the minimum value of the temperature scal...

Page 12: ... to move to the next programming mode Caution 1 To program the dummy analog output press SEL key in the engineering mode screen until OutC appears in the sub display as shown in the right figure 2 By pressing key the most significant digit of the main display will blink 3 Use or key to program the desired value 4 To shift to the next lower digit press key The next lower digit will blink 5 Repeat t...

Page 13: ...ain display Sub display Peak hold Tb CONT MEM PEAK AL AH C F ε Sub display Main display Sample hold Tb CONT MEM PEAK AL AH C F ε Main display Sub display 4 Engineering mode 1 To select the hold function type press SEL key in the engineering mode screen until HoLd appears in the sub display as shown in the right figure 2 By pressing key no PEAk or SAMP will blink in the main display 3 Use or key to...

Page 14: ...play 1 After the selection of PEAk in 4 3 1 Hold function type selection press SEL key until HrSt selection screen of peak hold reset type appears in the sub display as shown in the right figure 2 By pressing key no in or Et will blink in the main display 3 Use or key to select the desired type 4 To activate this selection press ENT key Tb CONT MEM PEAK AL AH C F ε External reset Main display Sub ...

Page 15: ...y Sub display 4 Engineering mode 1 After the selection of in time reset press SEL key until r tim programming screen of peak hold reset time appears in the sub display as shown in the right figure 2 By pressing key the most significant digit of the main display will blink 3 Use or key to program the desired value 4 To shift to the next lower digit press key The next lower digit will blink 5 Repeat...

Page 16: ...ey to select the decimal point 4 To activate this selection press ENT key The default is 2 decimal point 2 Column Remarks 1 To select the contacts output item type press SEL key in the engineering mode screen until dout appears in the sub display as shown in the right figure 2 By pressing key no ALM or Err will blink in the main display 3 Use or key to select the desired type 4 To activate this se...

Page 17: ... the sub display as shown in the right figure 2 By pressing key no in or Et will blink in the main display 3 Use or key to select the desired type 4 To activate this selection press ENT key Tb CONT MEM PEAK AL AH C F ε Calculation disable Main display Sub display Tb CONT MEM PEAK AL AH C F ε External automatic Emissivity calculation by analog input Main display Sub display Tb CONT MEM PEAK AL AH C...

Page 18: ...t 1 35 0 9µm InG2 400 to 1200ºC Single color type measurement 1 55µm 500 to 1400ºC Single color type measurement 1 55µm 350 to 900ºC Single color type measurement 1 55µm InG1 400 to 3100ºC Single color type measurement 1 35µm 500 to 3500ºC Single color type measurement 1 35µm 350 to 2000ºC Single color type measurement 1 35µm Si 800 to 3100ºC Single color type measurement 0 9µm 1000 to 3500ºC Sing...

Page 19: ...able range is 0 to 6280ºC The default is the minimum temperature value of the measuring range Remarks 1 Press SEL key in the zero side adjustments screen to enter span side adjustments SPan appears in the sub display as shown in the right figure 2 By pressing key the most significant digit of the main display will blink 3 Use or key to program the desired value 4 To shift to the next lower digit p...

Page 20: ... the 1 35 0 9µm after 4 9 1 1 Zero span adjustments execution selection 4 9 1 2 Zero side adjustments and 4 9 1 3 Span side adjustments 1 Select measuring wavelength inG2 in the 4 8 Measuring wavelength selection 2 Execute the zero span adjustments on the side of the 1 55µm after 4 9 1 1 Zero span adjustments execution selection 4 9 1 2 Zero side adjustments and 4 9 1 3 Span side adjustments 3 Nex...

Page 21: ...f 3 and 4 up to the least significant digit for programming 4 digits 6 To activate this programming press ENT key 1 To enter the zero side data display setting by furnace temperature by a reference radiation thermometer press SEL key in the 4 9 3 2 Zero side data display setting by IR CA series screen SEt Z appears in the sub display as shown in the right figure Press SEL key in the case of only t...

Page 22: ... next lower digit will blink 5 Repeat the above procedure of 3 and 4 up to the least significant digit for programming 4 digits 6 To activate this programming press ENT key 1 To enter the zero side data display setting by furnace temperature by a reference radiation thermometer press SEL key in the 4 9 3 4 Span side data display setting by IR CA series screen SEt S appears in the sub display as sh...

Page 23: ...g SEL key CAnc Canceling programmed desired value of 4 9 3 2 to 4 9 3 5 and exiting from this mode Data returns before the correction By pressing SEL key here it enters into 4 10 Temperature display Analog output correction programming screen Go Executing the zero span adjustments by correction command 4 Engineering mode 1 To enter the Selection of for correction after the zero span adjustments pr...

Page 24: ...creen until CP n appears in the sub display as shown in the right figure 2 By pressing key the least significant digit of the main display will blink 3 Use or key to program the desired value 4 To activate this programming press ENT key Main display Sub display The numeric figure in the main display is available from 0 to 9 However the programmable range is 0 and 2 to 8 The default value is 0 Rema...

Page 25: ...he next lower digit press key The next lower digit will blink 5 Repeat the above procedure of 3 and 4 up to the least significant digit for programming 4 digits 6 To activate this programming press ENT key Tb CONT MEM PEAK AL AH C F ε Main display Sub display The programmable range is 0 to 6280ºC The default is 0ºC Remarks 1 After the entering of the data before the temperature display and the ana...

Page 26: ...ction programming items cont Confirm and modify programmed desired value again By pressing SEL key here it enters into 4 10 2 Programming of temperature display and analog output correction data No screen For modifying the desired value modify the programmed value in reference to 4 10 2 up 4 10 4 Press SEL key in the case of only the confirmation of data The screen changes from 4 10 2 to 4 10 4 at...

Page 27: ...n the right figure 2 By pressing key no ErMt or EAut will blink in the main display 3 Use or key to select the desired type 4 To activate this selection press ENT key Analog input items no Analog input disable ErMt For remote emissivity programming EAut For automatic emissivity calculation Tb CONT MEM PEAK AL AH C F ε Analog input disable Main display Sub display Tb CONT MEM PEAK AL AH C F ε Remot...

Page 28: ...ng press SEL key in the engineering mode screen until AinH appears in the sub display as shown in the right figure 8 Repeat the above procedure from 2 to 6 Tb CONT MEM PEAK AL AH C F ε Minimum value of emisivity scaling Main display Sub display AL AH C F ε Tb CONT MEM PEAK Maximum value ofemisivity scaling Main display Sub display The programmable range is 1 999 to 0 001 The defaults are AinL Mini...

Page 29: ...e or key to program the desired value 4 To shift to the next lower digit press key The next lower digit will blink 5 Repeat the above procedure of 3 and 4 up to the least significant digit for programming 4 digits 6 To activate this programming press ENT key 7 To program the maximum temperature value for the scaling press SEL key in the engineering mode screen until AinH appears in the sub display...

Page 30: ...s available from 1 to 99 However the programmable range is 1 to 32 The default value is 1 Remarks 1 To select the communications speed press SEL key in the engineering mode screen until SPd appears in the sub display as shown in the right figure 2 By pressing key 4 8 9 6 or 19 2 will blink in the main display 3 Use or key to select the desired type 4 To activate this selection press ENT key Tb CON...

Page 31: ...hat Lock is displayed in the sub display 4 Lock will disappear and the screen will return to the measurement mode screen Tb CONT MEM PEAK AL AH C F ε Main display Sub display Keylock 4 Engineering mode 1 To select the communications protocol mode press SEL key in the engineering mode screen until Prot appears in the sub display as shown in the right figure 2 By pressing key rtu or PriV will blink ...

Page 32: ...played when the measured temperature is higher than the maximum value of measuring range 20 ºC uFL is displayed when the measured temperature is lower than the minimum value of measuring range 20 ºC 5 3 Under flow display for two color type radiation thermometer only CLP CLP is displayed when the measured temperature using two color type radiation thermometer is lower than minimum value of measuri...

Page 33: ...e engineering mode Ref 5 Engineering mode The screen automatically returns to the measurement mode screen when the programming up to 4 10 Analog output correction programming is completed except the thermometer with optional specifications or if any key is not pressed for 1minute Key lock The key lock is used not to change the parameters being programmed By pressing SEL and ENT keys simultaneously...

Page 34: ...ow temperature alarm Err Self diagnostic abnormal ALM 4 6 Automatic emissivity calculation E Aut no Calculation disable in Internal Key Et External Analog input no 4 7 Measuring wevelength selection dEtc 2col 1col 2c2 2c1 InG2 InG1 Si Refer to Measurment wavelength table 2col 4 8 Zero span adjustments CAL no Not execution yES Execution 4 9 Zero span adjustments 1 ZEro SPAn 0 to 6280ºC Minimum and ...

Page 35: ...nt Confirm and modify programmed desired value again CAnc Canceling programmed desired value and exiting from this mode Go Executing the temperature display and analog output correction by correction command 0 1 0ºC 0ºC cont 4 10 1 4 10 2 4 10 3 4 10 4 4 10 5 Analog input programming Ain no Analog input disable ErMt Remote emissivity programming EAut Automatic emissivity calculaiton no 4 11 1 Emis...

Page 36: ...O ITABAS HI KU TOKYO 173 8632 Telephone 81 3 3956 2171 Facsimile 81 3 3956 0915 Web site http www chino co jp INE 454 P2 Jan 05 IR CAQ Series High function type infrared radiation thermometer IR CAQ Printed in Japan ...

Reviews: