25
25
10A AC+DC
Warning the sum of AC + DC must never exceed the 6 A or 10 A range.
Range
Resolution
Specified
measurement
range
DC
Uncertainty
(
±
)
AC Uncertainty
(
±
)
Crest factor
Voltage drop
Protection
6 A
0.001 A
0.6 to 6 A 1.2 % R + 10 D
40 Hz to 1 kHz
1.2 % R + 5 D
2.8 at 5 A
0.05 V / mA
Fast fuse
10 A (or 11 A)
/1000 V
10 A / 20 A*
0.01 A
0.6 to 20.00 A 1 % R + 10 D
40 Hz to 1 kHz
1 % R + 3 D
3.7 at 8 A
The display shows "OL" in excess of 19.99 A. A beep sounds in excess of 10 A, 20 A for max 30s with a 5 min pause).
(*)
Admissible overload: 10 to 20 A for 30 s. max. with a pause of 5 min between 2 meas. Ambient temp. 35° C max.
- Secondary measurements and displays. Frequency (AC coupling): Fmax
≤
1 kHz, min max
4.2.13 Max/Min
Add 0.2 % R + 2 D to obtain the accuracy corresponding to the function and the range.
Max/Min capture time: 100 ms approx.