4.6 Compliance with international standards
Electrical safety
Application of safety rules in compliance with standards NF EN 61010-1 +
NF EN 61010-2-030
1000V, CAT III - 600V CAT IV, pollution level 2, double insulation.
Electromagnetic compatibility
Compliance with the NF EN 61326-1 and NF EN 61326-2-2 standards
Emission
:
class B
Immunity
:
Electrical discharge:
4 kV on contact, aptitude criterion B;
8 kV in the air, aptitude criterion B
Resistance to radiated fields: 10 V/m, aptitude criterion B
Resistance to rapid transients: 1 kV, aptitude criterion B
Conducted disturbances:
10 V/m, aptitude criterion A
Mechanical strength
Free fall:
1 m (tested in compliance with the
IEC 68-2-32 standard)
Shocks:
0.5 J (tested in compliance with the
IEC 68-2-27 standard)
Ingress protection
IP 54
in compliance with standard NF EN 60529