4. SPECIFICATIONS
4.1 Reference conditions
Influencing quantities:
Reference conditions
Temperature
23° C ± 5° C
Relative humidity
45 % to 75 %
Power supply voltage
9 V ± 1 V
Frequency domain of the applied signal
40 Hz to 1 kHz
Absence of electric field
NB: below, the accuracy is given as X % of the reading (R) ± Y counts (D). When the frequency exceeds 1 kHz, apply the
formula X % R + Y % x [F (kHz) - 1] R ± D
with F in kHz
.
4.2 Specifications of the reference conditions
Uncertainties are given in:
X % of the reading (R) ± Y counts (D).
When the frequency exceeds 1 kHz, apply the formula given in the tables
X % R + Y % x [F (kHz) - 0.4] R ± D.
where:
- Reading
"R",
- Digit
"D", the measurement count equivalent to the resolution of the display range,
- Frequency
"F" in kHz.
4.2.1 DC voltages
60 mV range
: The measurement of high currents or over a long period can cause certain components to heat. In this case,
a certain waiting time is needed to return to the specified metrological conditions. It is possible to check whether the offset
has returned to an acceptable value by short-circuiting the "+" and COM terminals. A reading of < 5 D should be obtained.
Display range
60 mV
1)
600 mV
6 V
60 V
600 V
1000 V
2)
Specified measurement
range
0 - 60.00 mV
0 - 600.0 mV 0 - 6.000 V 0 - 60.00 V
0 - 600.0 V
0 - 1000.0 V
Uncertainty (
±
)
0.5 % R + 5 D 0.5 % R + 2 D
0.09 % R + 2 D
Resolution
0.01 mV
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
1)
This range can only be accessed using the
key. Input impedance
≈
10.6 M
Ω
// 50 pF
2)
The display shows "+OL" in excess of +1050 V and "-OL" in excess of -1050 V.