60
MEASUREMENT MODE
Whatever the measurement mode, the sampling time is always 250µs. The table below
summarizes the measurement times of the various modes described in this chapter.
Measurement mode
Symbol
Digital measurement
time
Normal measurement
400 ms
Recording
(of MIN, MAX and AVG)
RECORD
400 ms
Smooth measurement
SMOOTH
4 s
Recording in smoothed values
RECORD SMOOTH
4 s
Peak measurement
PEAK
100 ms
Recording in peak values
RECORD PEAK
1 ms
HOLD
SMOOTH
SMOOTH
The time required for updating analogue measurements shown by the bargraph is always 20ms.
HOLD THE DIGITAL DISPLAY
One press on allows you to hold the digital display on the last measurement dis-
played whilst the analogue display continues to indicate the instantaneous value of the
measurement. The display indicates
HOLD
. Pressing again resets the instrument
to instantaneous measurement mode and
HOLD
disappears from the display.
SMOOTH MEASUREMENTS
A first press on triggers smoothing of the measurement (
SMOOTH
displayed).
The digital value shown is then the result of a sliding average calculated over the last 10
measurements (i.e. approx 4 seconds).
The bargraph still shows the instantaneous measurement. A second press on
cancels the filter and the
SMOOTH
symbol disappears.
The smoothed measurement value is constantly calculated. This allows you to obtain the
result of the filtering as soon as the button is pressed.
MEASUREMENT OF "PEAK" VALUES
The
PEAK
function allows you to make measurements with an acquisition speed of 1ms for
peak measurements.
HOLD