MEMORY MODE
It is possible to enter measurements in the memory when the selector switch is on one of
the two measurement positions: V/m (A/m) or µW/cm
≤
.
The memory can be accessed and read when the switch is set to MR.
MANUAL MEMORY
A single press on allows you to store all the parameters of the measurement that
are present, in the measurement memory, when the command is sent :
<Time> <Filter> <Measurement> <Unit>
< Time >
contains the time on the clock when the command is sent in HH MM
< Filter >
may contain
SMOOTH
or
PEAK
< Measurement >
contains the digital measurement displayed
< Unit >
contains the symbol for the measurement unit
Note :
The
HOLD
function, that may be active when the data is placed in the memory, is not held
in the memory.
When is pressed, the
MEM
symbol is displayed for one second and the digital
display flashes once to indicate the storage operation.
If you continue to press the button, the digital measurement is replaced by the number of the
box for the address in the memory (which has just been allocated), then the current memory
appears.
If the measurement memory is full the
MEM
symbol flashes for 2 seconds, the error beep
(button inoperative) is emitted and the value is not stored in memory. The maximum address,
1920, is displayed.
AUTOMATIC MEMORISATION
The C.A 43 Fieldmeter can be used for site monitoring. With a
∆
t session programmed from
1 minute to 24 hours, the min, max and average values corresponding to each session are
memorised automatically.
Firstly, set the session (see
∆
t programming). Then press to trigger automatic
memorisation of the MIN, MAX and AVG recordings at the selected interval (the
RECORD
and
MEM
symbols are displayed).
MEM
MIN MAX
MEM