84
Distortion magnitude
Limit of range
Magnitude distorted
MAX variation
Ambient
0 to 50
∞
C
all magnitudes
0.3 % /
∞
C
temperature
of the reading
± 0.5V/m per 10
∞
C
Humidity
10 to 90%
all
< 0.5 V/m
no condensation
magnitudes
Power supply
7.5 to 11 V
all magnitudes
0.05 %/V
Frequency of field
20 to 500 MHz
all magnitudes
± 1 dB
1 MHz to 1 GHz
± 1.5 dB
1 MHz to 2.5 GHz
± 2 dB
Level of field
0.1 to 10 V/m
± 0.5 V/m
0.1 to 100 V/m
all magnitudes
± 1 dB
0.1 to 200 V/m
± 2 dB
Frequency from
all magnitudes
± 1dB
Interchangeability
1 MHz to 2.5GHz
of sensors
Level of field
all magnitudes
± 0.5dB
from 0.1 to 200V/m
All distortion
from 0 to 50
∞
C
Threshold of
± 0.2V/m
magnitudes
from 10 to 90% RH
alarm detection
of the programmed
power supply from
value
7.5 to 10V
frequency from
1 MHz to 2.5GHz
level from
0.1 to 200V/m
■
Variation in the operating range
■
Adherence to standards
Class III instrument IEC 1010.
Electrostatic discharge (IEC 801-2 and IEC 1000-4-2)
Class of severity :
- level 2 (4 kV) no destruction of build components, but change of function that can be reset
by a new instruction.
- level 4 (15 kV) non destructive.
Radiated electrical fields (EN 55081-2 class B)
Protection from electromagnetic fields in accordance with standard EN 55082-2 up to 200V/m.
Class of severity: level 4 (200 V/m)