90
Coefficients of table 04 probe EF2 first sensitivity :
N
∞
straight line
Start
End
Coeff. a
Coeff. b
1
0 ct
27 cts
5.925 10e-2
0
2
27 cts
143 cts
1.207 10e-2
1.274
3
143 cts
572 cts
6.993 10e-3
2.000
4
572 cts
2544 cts
3.651 10e-3
3.911
5
2544 cts
8512 cts
1.776 10e-3
8.681
6
8512 cts
180224 cts
1.025 10e-3
15.07
Coefficients of table 05 probe EF2 second sensitivity :
N
∞
straight line
Start
End
Coeff. a
Coeff. b
1
0 ct
27 cts
5.925 10e-2
0
2
27 cts
143 cts
1.207 10e-2
1.274
3
143 cts
572 cts
7.459 10e-3
1.933
4
572 cts
2048 cts
4.268 10e-3
3.758
5
2048 cts
8000 cts
1.889 10e-3
8.611
6
8000 cts
175104 cts
1.053 10e-3
15.37
As for the example taken before, if the probe code is 227, use table 02.
The value of the measurement in points is 2802.4. It falls in the extreme values of the fifth
slope, thus the coefficients to take to linearise the measurement are :
a = 0.001893 and
b = 7.300
The time separating two readouts is limited by the rate of the instrument which is 20ms for a
rapid measurement.