O2X
DUO
- 17
Example of a handrail
when no handrail is
available.
•
Disassemble the plastic Rod guide and
place around the telescopic rod.
•
Use the SS screws on bracket to tighten the
bracket to the rod.
•
Insert the PVC bracket guide with the telescopic rod
into the SS mounting bracket. Make sure that the
bracket guide tracks are properly seated in the bracket.
•
Fasten the safety-locking clamp.
•
Check that the SS mounting bracket is safely fixed to
the handrail for the spring to work the way it is
intended.
•
To make it easier to inspect and service the
instrument, wind up a couple of meters of the cable
and fasten it on the rail with attached cable tie.
•
Hint on pick up: let the rod hang freely and lift it
straight up before you swing it over the rail.
•
Pull the M12 cable and flush hose through the sensor holder and rod.
•
Connect the telescoping fiberglass rod to the sensor with the two piece black plastic adapter.
•
The adapter halves should be tightened until snug, which will leave about 1-2 mm (1/16") gap.
NOTE!
The gap is required so the water can drain from the rod and sensor holder.
•
Adjust the length of the telescopic sensor rod as necessary by twisting the 2 nuts while holding
the rod.
NOTE!
Do not extend the rod sections beyond the black lines. This could lead to rod damage.