50
5. Maintenance
5.1 Changing the HEPA filter
1.
2.
3.
4.
5.
6.
7.
Turn off BIO X and disconnect the power cable.
Remove the Clean Chamber (tm) unit by reaching into the printchamber
and pushing up at the HEPA filter until the Clean Chamber (tm) unit
releases from the rest of BIO X.
Disconnect the cable powering the fans (1).
Put on a pair of gloves and sterilize them using >70 % ethanol.
Remove the four screws holding the HEPA filter in place (2).
Replace the HEPA filter with a clean one.
Fasten the clean HEPA filter with the four screws and plug in the cable
powering the fans.
Remount the Clean Chamber (tm) unit in the top of BIO X.
The HEPA filter is part of CELLINK’s patented Clean Chamber (tm) technology,
ensuring a sterile printchamber for your research. To ensure this, the HEPA filter must
be changed after 160 hours of runtime. Simply follow the instructions below to do this: