![Celera Motion MicroE Mercury II 6000 Series Installation Manual And Reference Manual Download Page 15](http://html.mh-extra.com/html/celera-motion/microe-mercury-ii-6000-series/microe-mercury-ii-6000-series_installation-manual-and-reference-manual_2579295015.webp)
Sensor Alignment and Calibration
IM-1004 | REV181127
Page 13
©2018 Celera Motion
Mercury II 6000 Series Encoders
Sensor Installation Manual and Reference Guide
5.2
Using the Software
Use the following instructions to perform sensor alignment and calibration using the Alignment
Tool and the SmartPrecision II Software.
5.2.1 Connect the Alignment Tool and Encoder
Step Action
1. Verify that the Alignment Tool Adapter/SmartPrecision II Alignment Tool is not powered on. The
Power/Cal
. LED should be off.
2. Connect the MII6000 encoder and Ethernet cable as shown below.
3. Power up the Alignment Tool by plugging in 5 VDC to the Alignment Tool Adapter.
Result:
The
Power/Cal.
LED illuminates a steady green. The sensor is initialized after an
approximate 10 second delay.
5 VDC Power
Alignment Tool Adapter
Power/Cal
. LED
Power/Cal
. LED
Alignment Tool
Encoder
Ethernet Cable
Sensor Connector
Computer
(Windows or
other OS)