Sensor Alignment and Calibration
IM-1004 | REV181127
Page 11
©2018 Celera Motion
Mercury II 6000 Series Encoders
Sensor Installation Manual and Reference Guide
Linear Scales or Rotary Scales used in applications less than 360°
Step Action
1.
To start calibration, press and hold the
Cal.
button for about two seconds until the
Power/Cal.
LED starts blinking slowly.
2.
Move the sensor 50 mm to perform
Gain/Offset/Phase calibration
. Move the sensor back and
forth if scale has <50 mm of measuring length.
Result
: After calibration both limit LEDs come on steady.
3.
Move the sensor to an area of the scale away from the index and limit markers. Press the
Cal.
button once quickly.
Result
: The
Power/Cal.
LED will start blinking quickly.
4.
Move the sensor over the index up to 20 passes (one pass is a single cycle back and forth).
Result
: The Left Limit LED will start blinking quickly.
Note
: If the sensor is positioned over the left limit marker, the Left Limit LED will come on
steady.
5.
Move the sensor over the left limit marker and press the
Cal.
button once quickly.
Result
: The Right Limit LED will start blinking.
Note
: if the sensor is positioned over the right limit marker, the Right Limit LED will come on
steady.
6.
Move the sensor over the right limit marker and press the
Cal.
button once quickly.
Result
: All LEDs will flash together twice to indicate that setup is completed.
The encoder is now ready for connection to the controller for use in servo control.
7.
When calibration is complete, go to
Section 6.0 Alignment Verification with Connector
LEDs
to visually verify calibration using the LED indicators on the MII6000 sensor’s connector.
Note:
To skip any portion of this calibration and move to the next procedure, press and hold the
Cal.
button for two seconds.
Power/Cal. LED
Cal.
Pushbutton
Limit Switches LEDs