Sensor Alignment and Calibration
IM-1004 | REV181127
Page 12
©2018 Celera Motion
Mercury II 6000 Series Encoders
Sensor Installation Manual and Reference Guide
Rotary Scales used in applications greater than 360° without Limit Markers
Step Action
1. To start calibration, press and hold the
Cal.
button for about 10 seconds: until the three
Signal
LEDs blink twice to indicate that rotary calibration has been activated.
2. Move the sensor 50 mm to perform
Gain/Offset /Phase calibration
. After calibration both limit
LED’s come on steady. Move the sensor back and forth if your scale has a circumference of
<50mm.
Result
: After calibration both limit LEDs come on steady.
3. Move the sensor to an area of the scale away from the index marker. Press the
Cal.
button
once quickly.
Result
: The
Power/Cal.
LED will start blinking quickly.
4. Move the sensor over the index up to 20 passes (one pass is a single cycle back and forth).
Result
: The
Power/Cal.
LED and both limit LEDs will start to blink in sets of three.
5. Move the sensor away from the index and press the
Cal
. button once quickly.
Result
: The
Power/Cal.
LED and both limit LED’s will start to blink in sets of two.
6. Move the sensor over the index once.
Result
: The LEDs will change to
Power/Cal.
LED and both limit LEDs blinking just once.
7. Make a full revolution of the rotary scale in order to go over the index again in the same
direction. The two passes over the index must be at least 1000 20 µm fringes apart (equivalent
of 20 mm linear travel), if they are not, the Alignment Tool will wait for another pass that is 1000
fringes from the first.
Result
: The encoder is now ready for connection to the controller for use in servo control.
8. When calibration is complete, go to
Section 6.0 Alignment Verification with Connector
LEDs t
o visually verify calibration using the LED indicators on the MII6000 sensor’s connector.
Note:
To skip any portion of` this calibration and move on to the next procedure, press and hold
the
Cal
. button for two seconds.
Signal LEDs (3)
Cal.
Pushbutton
Power/Cal. LED