Sensor Alignment and Calibration
IM-1004 | REV181127
Page 10
©2018 Celera Motion
Mercury II 6000 Series Encoders
Sensor Installation Manual and Reference Guide
5.1.2 Sensor Calibration
Perform calibration when installing the sensor for first time, or if the sensor is subsequently
remounted, or the scale is replaced. This section contains instructions for calibrating the following:
Linear Scales or Rotary Scales used in applications less than 360°
Rotary Scales used in applications greater than 360° without Limit Markers
Confirm Settings
Confirm the following before calibration:
Always perform calibration of the left and right limits while
Limit Polarity
is set to
Limits
Normal
mode. See the Left and Right Limits Settings in the
Calibration and Align
tab of
the SmartPrecision II Software in
Section 5.2.3 Sensor Alignment and Calibration
.
Select correct Reflective or Non-Reflective Grating Type on
Status and Setup
tab.
Note:
Perform all procedures below at
≤
1 meter/second relative motion between the sensor and
the scale.