10 APPENDIX A- Technical references
[15x5]
HiRes
CTDI
100 (centre)
12,1
18,1
-
24,8
-
CTDI
100 (0°)
14,4
23,2
-
29,4
-
CTDI
100 (90°)
14,0
20,9
-
28,6
-
CTDI
100 (180°)
12,0
18,0
-
24,5
-
CTDI
100 (270°)
13,2
19,5
-
26,9
-
CTDI
100 (periph.)
13,4
20,4
-
27,4
-
[12x8]
HiRes
CTDI
100 (centre)
10,1
15,1
-
20,5
-
CTDI
100 (0°)
11,6
17,4
-
23,5
-
CTDI
100 (90°)
10,3
15,7
-
21,0
-
CTDI
100 (180°)
8,4
12,8
-
17,1
-
CTDI
100 (270°)
9,7
14,7
-
19,8
-
CTDI
100 (periph.)
10,0
15,1
-
20,4
-
[8x8]
HiRes
CTDI
100 (centre)
9,5
14,2
-
19,2
-
CTDI
100 (0°)
8,9
13,6
-
18,1
-
CTDI
100 (90°)
7,9
11,8
-
16,1
-
CTDI
100 (180°)
6,6
9,8
-
13,3
-
CTDI
100 (270°)
7,6
11,5
-
15,5
-
CTDI
100 (periph.)
7,7
11,7
-
15,8
-
[6x6]
HiRes
CTDI
100 (centre)
10,5
12,8
-
21,3
-
CTDI
100 (0°)
8,7
10,7
-
17,6
-
CTDI
100 (90°)
7,5
9,1
-
15,1
-
CTDI
100 (180°)
6,3
7,7
-
12,8
-
CTDI
100 (270°)
7,4
9,1
-
15,1
-
CTDI
100 (periph.)
7,5
9,2
-
15,2
-
Date: 2013-07-11
20
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 108.90mAs - Enh. Scan Std Dose: 110kV 147.01mAs
21
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 108.90mAs - Enh. Scan Std Dose: 110kV 147.01mAs
22
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 108.90mAs - Enh. Scan Std Dose: 110kV 147.01mAs
23
Loading factors: automaticamente impostati dal software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 89.84mAs - Enh. Scan Std Dose: 110kV 147.01mAs
NewTom 5G
– User Manual
10-15
Summary of Contents for NewTom 5G
Page 1: ...NewTom 5G User Manual 0051 97050196 Rev 5 28 04 2017 EN...
Page 80: ...10 APPENDIX A Technical references NewTom 5G User Manual 10 7...
Page 81: ...10 APPENDIX A Technical references 10 8 NewTom 5G User Manual...
Page 82: ...10 APPENDIX A Technical references NewTom 5G User Manual 10 9...
Page 83: ...10 APPENDIX A Technical references 10 10 NewTom 5G User Manual...