10 APPENDIX A- Technical references
10.7 Dose declaration
CTDI
100
table declared for operative modalities of the device
Values are expressed in mGy. Tolerance is
30%. Values measured according to IEC 60601-2-44: 2009 Par 203.109.1
FOV
Eco Scan
Standard Dose
Regular Scan
Standard Dose
Regular Scan
Boosted Dose
Enhanced Scan
Standard Dose
Enhanced Scan
Boosted Dose
[18x16]
CTDI
100 (centre)
2,9
4,2
8,6
5,8
11,5
CTDI
100 (0°)
3,4
5,1
10,1
6,7
13,1
CTDI
100 (90°)
3,2
4,8
9,6
6,4
12,8
CTDI
100 (180°)
2,8
4,1
8,5
5,6
11,2
CTDI
100 (270°)
3,0
4,7
9,2
6,1
12,1
CTDI
100 (periph.)
3,1
4,7
9,4
6,2
12,3
[15x22e]
CTDI
100 (centre)
3,9
5,7
9,5
-
-
CTDI
100 (0°)
4,9
7,2
11,6
-
-
CTDI
100 (90°)
5,0
7,4
10,9
-
-
CTDI
100 (180°)
3,5
5,1
8,7
-
-
CTDI
100 (270°)
3,9
6,0
10,0
-
-
CTDI
100 (periph.)
4,3
6,3
10,3
-
-
[15x12]
CTDI
100 (centre)
2,9
4,2
8,7
5,8
10,7
CTDI
100 (0°)
3,6
5,1
10,1
7,1
13,7
CTDI
100 (90°)
3,4
5,0
10,2
6,8
14,0
CTDI
100 (180°)
2,8
4,1
8,4
5,6
11,4
CTDI
100 (270°)
3,0
4,5
9,0
6,0
12,6
CTDI
100 (periph.)
3,2
4,7
9,5
6,4
12,9
[12x8]
CTDI
100 (centre)
3,5
5,2
8,3
6,9
11,2
CTDI
100 (0°)
3,9
5,1
9,9
7,8
12,6
CTDI
100 (90°)
3,6
4,8
8,9
7,3
11,9
CTDI
100 (180°)
2,7
4,3
6,5
5,4
8,8
CTDI
100 (270°)
3,2
4,6
7,8
6,5
10,3
CTDI
100 (periph.)
3,4
4,7
8,3
6,7
10,9
[8x8]
CTDI
100 (centre)
3,3
5,2
8,2
6,9
11,0
CTDI
100 (0°)
3,1
4,7
7,7
6,4
9,6
CTDI
100 (90°)
2,8
4,2
6,8
5,7
9,0
CTDI
100 (180°)
2,2
3,3
5,4
4,5
7,6
CTDI
100 (270°)
2,5
4,0
6,3
5,2
8,7
CTDI
100 (periph.)
2,6
4,1
6,6
5,4
8,7
14
Measured by using “head phantom” according to IEC 60601-2-44: 2009 Par. 203.108
15
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 17.76mAs - Reg. Scan Std Dose: 110kV 26.35mAs - Reg. Scan Boosted Dose: 110kV 67.20mAs
- Enh. Scan Std Dose: 110kV 35.52mAs - Enh. Scan Boosted Dose: 110kV 87.64mAs
16
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 32,85mAs - Reg. Scan Std Dose: 110kV 48,84mAs - Reg. Scan Boosted Dose: 110kV 97,63mAs
17
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 19.07mAs - Reg. Scan Std Dose: 110kV 28.37mAs - Reg. Scan Boosted Dose: 110kV 71.89mAs
- Enh. Scan Std Dose: 110kV 38.14mAs - Enh. Scan Boosted Dose: 110kV 97.61mAs
18
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 23.14mAs - Reg. Scan Std Dose: 110kV 36.00mAs - Reg. Scan Boosted Dose: 110kV 64.21mAs
- Enh. Scan Std Dose: 110kV 46.28mAs - Enh. Scan Boosted Dose: 110kV 85.61mAs
19
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 25.30mAs - Reg. Scan Std Dose: 110kV 39.55mAs - Reg. Scan Boosted Dose: 110kV 71.25mAs
- Enh. Scan Std Dose: 110kV 50.60mAs - Enh. Scan Boosted Dose: 110kV 93.27mAs
10-14
NewTom 5G
– User Manual
Summary of Contents for NewTom 5G
Page 1: ...NewTom 5G User Manual 0051 97050196 Rev 5 28 04 2017 EN...
Page 80: ...10 APPENDIX A Technical references NewTom 5G User Manual 10 7...
Page 81: ...10 APPENDIX A Technical references 10 8 NewTom 5G User Manual...
Page 82: ...10 APPENDIX A Technical references NewTom 5G User Manual 10 9...
Page 83: ...10 APPENDIX A Technical references 10 10 NewTom 5G User Manual...