6 Scanning
6.2.1.2 Denture positioning with patient table with stretcher
1) Make sure that the stretcher is completely out of the gantry and locked by the handle positioned
at the side of the table console.
2) Press the
P2
button on the console in order to reset the patient table to the default position
(facilitated uphill position)
3) Remove the head support and press
P1
button on the console by moving the patient table to “exam
preparation” position.
4) Unlock the stretcher and slide it into the gantry. Then relock the stretcher.
5) Insert the denture in the provided slit of the denture support, and position the support on the table
carbon fiber board as shown in the following picture. Pay attention to do not reverse the denture
(must be in the same position as in the patient mouth).
6-14
NewTom 5G
– User Manual