9 IEC61223: Acceptance Test
CTDI
FREE AIR
CTDIvol
[
15x22e
]
other FOVs
CTDI
100
w:
[
18x16
]
[
15x22e
]
[
15x12
]
[
12x8
]
[
8x8
]
[
15x5
]
HiRes
[12x8] HiRes
[
8x8
]
HiRes
[
6x6
]
HiRes
N.A.
7.0 mGy
30%
Equal to CTDI
100
w
4.4
mGy
30%
6.4 mGy
30%
4.4
mGy
30%
3.8
mGy
30%
3.3
mGy
30%
10.2
mGy
30%
11.5
mGy
30%
9.6
mGy
30%
5.6
mGy
30%
Loading factors: automatically selected by the software.
Boosted Dose, Regular Scan
CTDI
FREE AIR
CTDIvol
[
15x22e
]
other FOVs
CTDI
100
w:
[
18x16
]
[
15x22e
]
[
15x12
]
[
12x8
]
[
8x8
]
N.A.
10.9 mGy
30%
Equal to CTDI
100
w
8.8
mGy
30%
10.0 mGy
30%
9.2
mGy
30%
6.2
mGy
30%
5.3
mGy
30%
Loading factors: automatically selected by the software.
Standard Dose, Enhanced Scan
CTDI
FREE AIR
CTDIvol
CTDI
100
w:
[
18x16
]
[
15x12
]
[
12x8
]
[
8x8
]
[
15x5
]
HiRes
[12x8] HiRes
[
8x8
]
HiRes
[
6x6
]
HiRes
N.A.
Equal to CTDI
100
w
5.8
mGy
30%
5.8
mGy
30%
5.0
mGy
30%
4.4
mGy
30%
13.5
mGy
30%
15.3
mGy
30%
12.8
mGy
30%
9.4
mGy
30%
Loading factors: automatically selected by the software.
Boosted Dose, Enhanced Scan
NewTom 5G
– User Manual
9-3
Summary of Contents for NewTom 5G
Page 1: ...NewTom 5G User Manual 0051 97050196 Rev 5 28 04 2017 EN...
Page 80: ...10 APPENDIX A Technical references NewTom 5G User Manual 10 7...
Page 81: ...10 APPENDIX A Technical references 10 8 NewTom 5G User Manual...
Page 82: ...10 APPENDIX A Technical references NewTom 5G User Manual 10 9...
Page 83: ...10 APPENDIX A Technical references 10 10 NewTom 5G User Manual...