NewTom 5G
series
–
Patient Table Service Procedures
2-3
To be used in case, despite the activation of "MOVE ALL ON", the leg rest doesn't move.
-
REF. BACKREST:
Search for the minimum back rest plate limit switch; move the back rest plate to
90°. To be used in case, despite the activation of "MOVE ALL ON", the back rest doesn't move.
-
REF.LIFT MOTOR:
Search for the minimum pantograph limit switch; low down the pantograph to
the minimum height. To be used in case, despite the activation of "MOVE ALL ON", the pantograph
rest doesn't move.
-
CONSOLLE RIGHT:
Store the console position as at the patient's right.
-
CONSOLLE LEFT:
Store the console position as at the patient's left.
-
DAILY REFER.:
Perform the daily reset procedure (same as the first movement sequence required
after the device has been turned on)
-
FULL REFER.:
Perform the complete reset procedure.
-
REF.UP-DOWN ALL [PW]:
Search for the minimum and maximum limit switch of back rest plate,
leg rest plate, pantograph. Can be run only with the table in the seat position and chair all out of
the gantry
(X = 0mm). ATTENTION: risk of collision between mechanical parts!
-
LOAD DEFAULT [PW]:
Restore the original manufacturer settings (without resetting the
calibrations)
-
GO PACKAGING [PW]:
Bring the table to the position for the packaging.
-
SET HORIZ POS [PW]:
Store in the EEPROM the current position of the back rest plate and consider
it as the "Horizontal Headrest" position (position in which the the back rest plate is moved at the
end of the “P1” button corresponding procedure).
-
SET SERIAL NUM [PW]:
Set the table serial number. The serial number can be adjusted by pressing
the “+” and “-” buttons.
-
XY CALIBRATION [PW]:
XY axis board potentiometer calibration. To be performed only after
replacement of the XY motor board and/or potentiometer. Can be run only with the table in the
seat position. (ATTENTION: risk of collision between mechanical parts!)
-
POTENT.CALIB. [PW]:
Pantograph potentiometer calibration: to be run only after replacement of
pantograph motor/potentiometer. Can be run only with the table in the seat position.
(ATTENTION: risk of collision between mechanical parts!)
-
XY LIMITS OFF [PW]:
Disable the software limit switches for the XY motor (only the electric limit
switches are enabled)
-
EXIT:
Exit from the ACTIONS page e return to the TOOLS page.
Once the desired action has been selected the monitor will prompt for the user confirmation.
CONFIRM page