 Patient Table Service Procedures 

 

3-10

 

The following table  includes error codes that could  appear on the console  display and possible  causes / 
solutions. 
 

Err. Code 

Description 

Possible causes / Solutions 

Power supply of Hanning boards < 26VDC 
 

40VDC  power  supply  of  Hanning  boards  may  be 
missing or may not be connected properly 

Power supply of Hanning boards > 43VDC 

40VDC  power  supply  of  Hanning  boards  not  be 
connected properly. 

Motor or Hanning pantograph board overheating 

Wait  2  minutes  and  try  again  the  movement.  If  the 
error  persists,  there  may  be  a  malfunction  of  the 
pantograph  motor  or  the  pantograph  is  not  free  to 
move. 

Overcurrent  peak  output  to  Hanning  pantograph 
board 

There may be a malfunction of the pantograph motor 

13 

Short-curcuit output to Hanning boards 

The  pantograph  motor  may  be  faulty,  or  incorrect 
wiring  of  the  motor  power  supply:  X102,  X103,  X104 
on one of the Hanning boards. 

16 

Enable  external  command  not  present  on  Hanning 
boards  

May be missing the enable jumper between pins 3 and 
2 of X208. 

17 

Hanning  pantograph  board  movement  sequence  time 
out 

The Hanning board may be faulty or not programmed 
properly or the pantograph motor is locked 

18 

Error  on  the  start  of  the  movement  of  the  Hanning 
pantograph board. 

The Hanning board may be faulty or not programmed 
properly 

128 

Internal timing error on one of the Hanning boards. 

The Hanning board may be faulty or not programmed 
properly 

129 

System error of Hanning boards 

The Hanning board may be faulty or not programmed 
properly 

131 

Reset for Hanning boards error 

The Hanning board may be faulty or not programmed 
properly 

132 

Reset for power supply change in Hanning boards 

40VDC  power  supply  of  Hanning  boards  may  have 
undergone rapid > 5 VDC 

133 

Software reset of Hanning boards 

The Hanning board may be faulty or not programmed 
properly 

140 

CRC error on parameters reading of hanning boards 

The Hanning board may be faulty or not programmed 
properly 

141 

Read error of Hanning board table 

The Hanning board may be faulty or not programmed 
properly 

142 

Read  error on parameters of Hanning boards 

The Hanning board may be faulty or not programmed 
properly 

150 

XY  axes  controller  board  doesn’t  send  regularly  the 
status message.

 

The board may be turned off, faulty or CANBUS is not 
working properly. 

151 

Hanning  pantograph  board  doesn’t  respond  correctly 
if cyclically interrogated

 

The board may be faulty, not powered or with wrong 
address. CAN may be faulty 

152 

The Hanning board that control the stretcher position 
and  handle  release  doesn't  respond  properly  if 
cyclically interrogated

 

The board may be faulty, not powered or with wrong 
address. CAN may be faulty 

210 ÷ 213 

Error during motion started by remote control   

The  system  can  not  reach  the  required  positions  or 
the movement is stopped before its completition 

253 

There  is  a  position  mismatch  between  the 
potentiometer and pantograph encoder 

It  may  not  have  been  carried  out  a  correct 
calibration,  the  Hanning  board  may  have  lost  some 
encoder  count  or  may  be  an  hardware  problem  to 
encoder, potentiometer or wiring. 

254 

Calibration  data  stored  in  the  console  are  not 
compatible  with  the  new  firmware  version  just 
installed and consequently restored 

The  console  board  may  have  been  replaced  or 
updated with new firmware without calibrations 

 

Summary of Contents for NewTom 5G Series

Page 1: ...97050267 Rev 3 28 10 2016 NewTom 5G series Patient Table Service Manual EN...

Page 2: ...269 EEC 90 270 EEC 90 394 EEC and 90 679 EEC concerning the improvement of the health and safety of workers in the workplace and with the essential requirements set forth by the Italian Legislative De...

Page 3: ...4 2 1 3 DATA DEBUG page 2 4 3 PATIENT TABLE WITH STRETCHER 3 1 3 1 SERVICE LEVEL PROCEDURES 3 1 3 1 1 Service Level Access 3 1 3 1 2 TOOLS page 3 2 3 1 2 1 ACTION page 3 2 3 1 2 2 LIMIT SWITCH page 3...

Page 4: ...please refer to the Service Manual document This manual is intended for trained personnel recognized by the manufacturer of the NewTom 5G series devices Prior to operating or servicing this device thi...

Page 5: ...ject and or person 2 1 1 Service Level Access From the INFO page hold the button and press the button The SERVICE Level modality will be enabled In order to disable the SERVICE Level modality repeat t...

Page 6: ...alue using the and buttons The value will be incremented decremented by 1 10 or 100 units depending on the pressure duration The value is reset every time the user scroll among different actions and i...

Page 7: ...anical parts LOAD DEFAULT PW Restore the original manufacturer settings without resetting the calibrations GO PACKAGING PW Bring the table to the position for the packaging SET HORIZ POS PW Store in t...

Page 8: ...gnals In the bottom right corner the software version is shown By pressing the MODE button is possible to return to the TOOLS page LIMIT SWITCH page X X Axis gantry in out Y Patient s left right Z Pan...

Page 9: ...th object and or person WARNING Every movement can be stopped by pressing the emergency button located under the console keypad only if the patient table is powered through the 5G series gantry 3 1 1...

Page 10: ...the F2 button POS page by pressing MODE button 3 1 2 1 ACTION page The ACTIONS page contain a scroll menu with a list of possible action to perform The F1 and F2 buttons allow to move among the list...

Page 11: ...he and buttons Change digit by the arrows keys NEW TABLE POT TYPE To setup only in case the small potentiometer cod 97643500 is mounted enabled by default Perform only for patient table with S N 0005...

Page 12: ...ress OK to confirm 5 Turn off the machine 6 Connect C6 connector 5G patient table J8 connector and X207 connector 7 Turn on the machine and perform the Pantograph potentiometer calibration by select P...

Page 13: ...RDS command from the service menu insert the password and press OK to confirm 12 Turn off the machine 13 Connect C6 connector 5G patient table J8 connector and X207 connector 14 Turn on the machine EX...

Page 14: ...completely out of the gantry switch Stretcher all the way out of the gantry UT Unlock table switch Stretcher lock A Anti pitch safety switch AntiPinch activated E Emergency button switch Emergency bu...

Page 15: ...k the pantograph by using BACK button In case of table installation console replacement or firmware update perform the following calibrations 1 Insert the device serial number following procedure SET...

Page 16: ...n case of potentiometer replacement or after adjustment of the limit switch position follow the next steps 1 Select the type of mounted potentiometer following procedure NEW TABLE POT TYPE OLD TABLE P...

Page 17: ...e quota 2 mm Longitudinal pantograph movement relative quota 200 210 mm Vertical pantograph movement relative quota 300 315 internal units The following table shows the stretcher pantograph the correc...

Page 18: ...boards error The Hanning board may be faulty or not programmed properly 132 Reset for power supply change in Hanning boards 40VDC power supply of Hanning boards may have undergone rapid 5 VDC 133 Sof...

Page 19: ...NewTom 5G series Patient Table Service Procedures 3 11 3 1 6 Hanning boards connections and interfaces 97661177 97661178...

Page 20: ...3 1 7 List of patient table electronic boards UPPER STRUCTURE BOARDS 1 97661178 97660848 TABLE BOARD 2 97660897 XY MOTORS CONTROL BOARD 3 97660888 MANAGEMENT BOARD PANTOGRAPH BOARDS 1 97660905 POWER S...

Page 21: ...NewTom 5G series Patient Table Service Procedures 3 13 CONSOLLE BOARD 97661095 CONSOLLE BOARD...

Page 22: ...NewTom 5G series Patient Table Service Procedures 3 14 3 1 8 Screens flow chart...

Page 23: ...emark of CEFLA s c All other products and brand names are registered trademarks or trademarks of their respective companies NEWTOM 5G is manufactured by CEFLA s c Phone 39 045 8202727 Fax 39 045 82030...

Reviews: