 Patient Table Service Procedures 

 

2-2

 

2.1.2 

 

TOOLS page 

 
In the 

TOOLS page

 is possible to select the  

 

“ACTIONS” page by pressing F1 button 

“LIMIT SWITCH” page by pressing the F2 button.  

“POS” page by pressing MODE button. 

 

 

TOOL page 

2.1.2.1

 

ACTIONS page 

 
The 

ACTIONS page

 contain a scroll menu with a list of possible action to perform. 

The  F1  and  F2  buttons  allow  to  move  among  the  list  of  actions  that  can  be  seleted  pressing  the  MODE 
button. 

 

ACTIONS page 

 
Some  actions  require  a  password  in  order  to  be  selected.  In  that  case  besides  the  description  of  the 
specific action a  4 digit value will appear. 
It possible to modify the value using the “+” and “-” buttons. The value will be incremented/decremented 
by  1,  10  or  100  units  depending  on  the  pressure  duration. The  value  is  reset  every  time  the  user  scroll 
among different actions and if necessary the correct password must be typed again. 
 
The 

PASSWORD

 value is 

9469.

 

 

 

ACTIONS page showing action with password request 

 

Pressing  the  MODE  button  with  a  wrong  password  will  result  in  showing  for  a  few  seconds  a  “

ACCESS 

DENIED!

”  message.   

 
The provided actions are: 
 

MOVE ALL ON:

 Allow to enable the single movement of all axis. After the command has been 

enabled the XY axis can be manually moved. Other movements (back rest plate, leg rest plate and 
pantograph) could  required the corresponding "referencing" signal, before to be manually moved. 

 

REF. LEGREST:

 Search for the minimum leg rest plate limit switch; move the leg rest plate to  90°. 

Summary of Contents for NewTom 5G Series

Page 1: ...97050267 Rev 3 28 10 2016 NewTom 5G series Patient Table Service Manual EN...

Page 2: ...269 EEC 90 270 EEC 90 394 EEC and 90 679 EEC concerning the improvement of the health and safety of workers in the workplace and with the essential requirements set forth by the Italian Legislative De...

Page 3: ...4 2 1 3 DATA DEBUG page 2 4 3 PATIENT TABLE WITH STRETCHER 3 1 3 1 SERVICE LEVEL PROCEDURES 3 1 3 1 1 Service Level Access 3 1 3 1 2 TOOLS page 3 2 3 1 2 1 ACTION page 3 2 3 1 2 2 LIMIT SWITCH page 3...

Page 4: ...please refer to the Service Manual document This manual is intended for trained personnel recognized by the manufacturer of the NewTom 5G series devices Prior to operating or servicing this device thi...

Page 5: ...ject and or person 2 1 1 Service Level Access From the INFO page hold the button and press the button The SERVICE Level modality will be enabled In order to disable the SERVICE Level modality repeat t...

Page 6: ...alue using the and buttons The value will be incremented decremented by 1 10 or 100 units depending on the pressure duration The value is reset every time the user scroll among different actions and i...

Page 7: ...anical parts LOAD DEFAULT PW Restore the original manufacturer settings without resetting the calibrations GO PACKAGING PW Bring the table to the position for the packaging SET HORIZ POS PW Store in t...

Page 8: ...gnals In the bottom right corner the software version is shown By pressing the MODE button is possible to return to the TOOLS page LIMIT SWITCH page X X Axis gantry in out Y Patient s left right Z Pan...

Page 9: ...th object and or person WARNING Every movement can be stopped by pressing the emergency button located under the console keypad only if the patient table is powered through the 5G series gantry 3 1 1...

Page 10: ...the F2 button POS page by pressing MODE button 3 1 2 1 ACTION page The ACTIONS page contain a scroll menu with a list of possible action to perform The F1 and F2 buttons allow to move among the list...

Page 11: ...he and buttons Change digit by the arrows keys NEW TABLE POT TYPE To setup only in case the small potentiometer cod 97643500 is mounted enabled by default Perform only for patient table with S N 0005...

Page 12: ...ress OK to confirm 5 Turn off the machine 6 Connect C6 connector 5G patient table J8 connector and X207 connector 7 Turn on the machine and perform the Pantograph potentiometer calibration by select P...

Page 13: ...RDS command from the service menu insert the password and press OK to confirm 12 Turn off the machine 13 Connect C6 connector 5G patient table J8 connector and X207 connector 14 Turn on the machine EX...

Page 14: ...completely out of the gantry switch Stretcher all the way out of the gantry UT Unlock table switch Stretcher lock A Anti pitch safety switch AntiPinch activated E Emergency button switch Emergency bu...

Page 15: ...k the pantograph by using BACK button In case of table installation console replacement or firmware update perform the following calibrations 1 Insert the device serial number following procedure SET...

Page 16: ...n case of potentiometer replacement or after adjustment of the limit switch position follow the next steps 1 Select the type of mounted potentiometer following procedure NEW TABLE POT TYPE OLD TABLE P...

Page 17: ...e quota 2 mm Longitudinal pantograph movement relative quota 200 210 mm Vertical pantograph movement relative quota 300 315 internal units The following table shows the stretcher pantograph the correc...

Page 18: ...boards error The Hanning board may be faulty or not programmed properly 132 Reset for power supply change in Hanning boards 40VDC power supply of Hanning boards may have undergone rapid 5 VDC 133 Sof...

Page 19: ...NewTom 5G series Patient Table Service Procedures 3 11 3 1 6 Hanning boards connections and interfaces 97661177 97661178...

Page 20: ...3 1 7 List of patient table electronic boards UPPER STRUCTURE BOARDS 1 97661178 97660848 TABLE BOARD 2 97660897 XY MOTORS CONTROL BOARD 3 97660888 MANAGEMENT BOARD PANTOGRAPH BOARDS 1 97660905 POWER S...

Page 21: ...NewTom 5G series Patient Table Service Procedures 3 13 CONSOLLE BOARD 97661095 CONSOLLE BOARD...

Page 22: ...NewTom 5G series Patient Table Service Procedures 3 14 3 1 8 Screens flow chart...

Page 23: ...emark of CEFLA s c All other products and brand names are registered trademarks or trademarks of their respective companies NEWTOM 5G is manufactured by CEFLA s c Phone 39 045 8202727 Fax 39 045 82030...

Reviews: