Patient Table Service Procedures
3-8
NEW TYPE
potentiometer (cod. 97643500)
mounted on patient table with S/N:
0005 , 0012 and subsequent
4)
Calibration of the stretcher potentiometer (following procedure
TABLE POT. CALIB)
5)
Calibration of the pantograph potentiometer (following procedure
PANT. FC BOTH
)
6)
XY Calibration (following procedure
XY CALIBRATION)
Once all the calibration steps has been carried out restart the device and verify the correct functionality.
In case of
potentiometer replacement
or after
adjustment of the limit switch position
follow the next
steps:
1)
Select the type of mounted potentiometer (following procedure
NEW TABLE POT. TYPE / OLD
TABLE POT. TYPE)
2)
Calibration of the stretcher potentiometer (following procedure
TABLE POT. CALIB)
3)
Calibration of the pantograph potentiometer (following procedure
PANT. FC BOTH
)
4)
XY Calibration (following procedure
XY CALIBRATION)
Once all the calibration steps has been carried out restart the device and verify the correct functionality.
In case of replacement of boards responsible for the pantograph movement or for the strectcher position
reading please contact service support.