NewTom 5G
series
–
Patient Table Service Procedures
3-7
3.1.4
Calibration sequence
WARNING:
In case of it is necessary to perform PANT. FC BOTH
calibration, if the patient table is mechanical connected with
the 5G series gantry, exit from TOOLS page, slide the
stretcher completely out of the gantry and completely draw
back the pantograph by using
BACK
button.
In case of
table installation
,
console replacement
or
firmware update
, perform the following
calibrations:
1)
Insert the device serial number (following procedure
SET SERIAL NUMBER
)
2)
Set the console position (following procedure
CONSOLLE RIGHT/ LEFT
)
Consolle
right
Consolle
left
3)
Select the right type of mounted potentiometer (following procedure
NEW TABLE POT. TYPE /
OLD TABLE POT. TYPE)
OLD TYPE
potentiometer (cod. 97643400)
mounted on patient table with S/N:
0001 ÷ 0004 , 0006 ÷ 0011 (included
)