NewTom 5G
series
–
Patient Table Service Procedures
2-1
2
Patient Table
2.1
Service Level Procedures
Special keys combination of the table console, allow to enable SERVICE level (“support modality”)
DANGER:
The usage of the SERVICE level by not authorized
personnel can result in mechanical damages of the
table and/or the scanner and injury of the user.
The SERVICE level must be used without any patient on
the table.
Every movement can be stopped by pressing the
emergency button located under the console keypad.
WARNING:
Pay attention while moving the patient's table
in order to avoid possible collision with object
and/or person.
2.1.1
Service Level Access
From the INFO page hold the
button and press the
button. The SERVICE Level modality will be
enabled.
In order to disable the SERVICE Level modality repeat the same sequence or turn off and on the device.
INFO page
Once the
SERVICE Level
is enabled on the
POS page
the
TOOLS page
and
DATA DEBUG page
will appear.
Each page can be selected by pressing the corresponding F1 or F2 buttons.
POS page with SERVICE Level
While in the
POS PAGE
it is possible to show some additional informations:
-
the “X” position will toggle to the “ K ” position when pressing “ + ” on the console
-
the “Pantograph Encoder” position will toggle to the “Pantograph Potentiometer” position when
pressing “ - ” on the consolle
X or K
Y
Z
LEGREST ENCODER
BACKREST ENCODER
PHANTOGRAPH ENCODER or
POTENTIOMETER VALUE