NewTom 5G
series
–
Patient Table Service Procedures
3-9
3.1.5
Troubleshooting
The following table shows the stretcher / pantograph the correct ranges verificabili from the "POS" page
(by pressing P1 button and with the stretcher completely extracted from the gantry)
Stretcher insertion longitudinal absolute quota
310 ÷ 320
[mm]
Stretcher transversal movement absolute quota
-5 ÷ 5
[mm]
Stretcher vertical movement absolute quota
735 ÷ 745
[mm]
Stretcher potentiometer relative quota
≤
2
[mm]
Longitudinal pantograph movement relative quota
200 ÷ 210
[mm]
Vertical pantograph movement relative quota
300 ÷ 315
[internal units]
The following table shows the stretcher / pantograph the correct ranges verificabili from the "POS" page
(by pressing P2 button and with the stretcher completely extracted from the gantry)
Stretcher insertion longitudinal absolute quota
220 ÷ 230
[mm]
Stretcher transversal movement absolute quota
-5 ÷ 5
[mm]
Stretcher vertical movement absolute quota
515 ÷ 525
[mm]
Stretcher potentiometer relative quota
≤
2
[mm]
Longitudinal pantograph movement relative quota
260 ÷ 270
[mm]
Vertical pantograph movement relative quota
47 ÷ 57
[internal units]