ALARMS
Alarm System Overview
The chamber light intensity and control system is equipped with an alarm system that
constantly monitors temperature, and humidity (on humidified models) to ensure the
user is notified if the cabinet goes into an alarm condition. Notification occurs via an
alarm pop-up window and a buzzer. Each alarm condition has been factory
programmed to minimize nuisance alarms while maximizing warning time. There is a 2
hour time delay after start-up and setpoint changes. To avoid nuisance alarms after a
routine door opening, an alarm condition must be present for 15 minutes (45 minutes for
humidity) before the operator is alerted. If the optional remote alarm contacts are
present, in an alarm condition, the dry contacts will change state.
The following alarm messages may be displayed:
•
Chamber temperature is higher than setpoint temperature
•
Chamber temperature is lower than setpoint temperature
•
Chamber humidity is higher than setpoint humidity
•
Chamber humidity is lower than setpoint humidity
•
Chamber CO
2
is higher than setpoint humidity
•
Chamber CO
2
is lower than setpoint humidity
•
Chamber light intensity is higher than setpoint intensity
•
Chamber light intensity is lower than setpoint intensity
•
Temperature sensor error
In the event an alarm occurs, the alarm indicator will appear on the status bar and an
audible alarm pop-up window will automatically appear.
The flashing
(Alarm) icon will appear on the status bar.
Alarm icon
Audible speaker
Summary of Contents for 7310-22
Page 1: ......
Page 115: ...7310 Series Operations Manual REV E 11 19 19 Page 115 of 131 ELECTRICAL SCHEMATICS ...
Page 116: ...7310 Series Operations Manual REV E 11 19 19 Page 116 of 131 ...
Page 117: ...7310 Series Operations Manual REV E 11 19 19 Page 117 of 131 ...
Page 118: ...7310 Series Operations Manual REV E 11 19 19 Page 118 of 131 ...
Page 119: ...7310 Series Operations Manual REV E 11 19 19 Page 119 of 131 ...
Page 120: ...7310 Series Operations Manual REV E 11 19 19 Page 120 of 131 ...
Page 121: ...7310 Series Operations Manual REV E 11 19 19 Page 121 of 131 ...
Page 122: ...7310 Series Operations Manual REV E 11 19 19 Page 122 of 131 ...
Page 123: ...7310 Series Operations Manual REV E 11 19 19 Page 123 of 131 ...
Page 124: ...7310 Series Operations Manual REV E 11 19 19 Page 124 of 131 ...