Calibrating Temperature
If temperature calibration is needed, perform either Method A or Method B.
Method A (geometric center)
Locate the reference instrument’s temperature sensor in close proximity to the cabinet’s
geometric center. Turn the lights ‘
off
’. Let the unit stabilize for at least 3 hours at the
temperature setpoint.
Method B (chamber sensor)
Locate the reference instrument’s temperature sensor behind the rear wall duct sheet
and adjacent the chamber’s existing temperature sensor (within 1 inch (25 mm)). See
temperature sensor location in the Equipment Overview section of this manual. Be sure
to re-install the rear wall access door panel that covers the sensors. For this method,
the lights may be ‘
on
’ or ‘
off
’. Let the unit stabilize for at least 3 hours (lights ‘
off
’) or 6
hours (lights ‘
on
’) at the temperature setpoint.
At the calibrate screen, press the
(Temperature Calibrate) button.
Temperature
Calibrate Button
In both temperature calibration methods, the
temperature sensor
must be
out of incident light
during the calibration process.
Summary of Contents for 7310-22
Page 1: ......
Page 115: ...7310 Series Operations Manual REV E 11 19 19 Page 115 of 131 ELECTRICAL SCHEMATICS ...
Page 116: ...7310 Series Operations Manual REV E 11 19 19 Page 116 of 131 ...
Page 117: ...7310 Series Operations Manual REV E 11 19 19 Page 117 of 131 ...
Page 118: ...7310 Series Operations Manual REV E 11 19 19 Page 118 of 131 ...
Page 119: ...7310 Series Operations Manual REV E 11 19 19 Page 119 of 131 ...
Page 120: ...7310 Series Operations Manual REV E 11 19 19 Page 120 of 131 ...
Page 121: ...7310 Series Operations Manual REV E 11 19 19 Page 121 of 131 ...
Page 122: ...7310 Series Operations Manual REV E 11 19 19 Page 122 of 131 ...
Page 123: ...7310 Series Operations Manual REV E 11 19 19 Page 123 of 131 ...
Page 124: ...7310 Series Operations Manual REV E 11 19 19 Page 124 of 131 ...