9
The subject should then inhale to TLC (Total Lung Capacity), lungs
full, then perform a ‘Valsalva’ manoeuvre, a forced exhalation
against the MicroRPM with as much effort as possible for as long as
possible (minimum 2 seconds).
The display will report the result, the maximum average expiratory
pressure sustained over a 1 second period of the test, in cmH2O.
Ideally, the subject should repeat this test 3 times to ascertain a
best value.
To repeat either the PImax or PEmax tests the MicroRPM must firstly
be returned to the ‘Off’ position.
_____
For more information on these procedures, and for interactive
presentations and video demonstrations, go to our dedicated
website:
WWW.RESPIRATORYMUSCLES.COM
____
Henrochech nv - Acacialaan 8 - 2630 Aartselaar - www.henrotech.be - [email protected]
All manuals and user guides at all-guides.com
Summary of Contents for MicroRPM
Page 6: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 20: ...All manuals and user guides at all guides com...
Page 35: ...31 3 2 4 6 8 2 9 1 SNIP MIP MEP OFF 3 1 7 4 5 All manuals and user guides at all guides com...
Page 60: ...All manuals and user guides at all guides com...
Page 64: ...59 1 P max OFF MIP MEP All manuals and user guides at all guides com...
Page 65: ...60 0 TLC 2 OFF PI max OFF MIP MEP RV 2 All manuals and user guides at all guides com...
Page 66: ...61 4 OFF SNIP All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 67: ...62 0 SNIP FRC OFF BAT MicroRPM PUMA RS 232 All manuals and user guides at all guides com...
Page 68: ...63 MicroRPM 15 CareFusion 65 All manuals and user guides at all guides com...
Page 69: ...64 MicroRPM SNIP 200 300 MicroRPM 3 MicroRPM All manuals and user guides at all guides com...
Page 75: ...Notes All manuals and user guides at all guides com...