13
Manometer
cm H O
2
0
2
1
3
6
4
5
8
7
9
Syringe
10
Connecting
line
Female
Luer
T Piece
Nasal plug
adapter
Slide the MicroRPM switch from ‘Off’ to ‘SNIP’ whilst applying no
pressure to the unit.
Gently fill the syringe until negative pressure of between 200 and
300cmH2O is read on the Water Manometer. Check that the reading
on the MicroRPM is within 3% of this reading.
Please Note:
Calibration can only be adjusted in a positive
direction, as the MicroRPM measures the highest peak value.
Therefore, if the reading was greater than that on the Water
Manometer, the calibration screw must be turned anti-clockwise
first, before calibration is attempted.
Henrochech nv - Acacialaan 8 - 2630 Aartselaar - www.henrotech.be - [email protected]
All manuals and user guides at all-guides.com
all-guides.com
Summary of Contents for MicroRPM
Page 6: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 20: ...All manuals and user guides at all guides com...
Page 35: ...31 3 2 4 6 8 2 9 1 SNIP MIP MEP OFF 3 1 7 4 5 All manuals and user guides at all guides com...
Page 60: ...All manuals and user guides at all guides com...
Page 64: ...59 1 P max OFF MIP MEP All manuals and user guides at all guides com...
Page 65: ...60 0 TLC 2 OFF PI max OFF MIP MEP RV 2 All manuals and user guides at all guides com...
Page 66: ...61 4 OFF SNIP All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 67: ...62 0 SNIP FRC OFF BAT MicroRPM PUMA RS 232 All manuals and user guides at all guides com...
Page 68: ...63 MicroRPM 15 CareFusion 65 All manuals and user guides at all guides com...
Page 69: ...64 MicroRPM SNIP 200 300 MicroRPM 3 MicroRPM All manuals and user guides at all guides com...
Page 75: ...Notes All manuals and user guides at all guides com...