26
Turn the respiratory pressure meter on to the SNIP position.
Very slowly fill the syringe until a negative pressure of between 200
and 300cm is obtained on the manometer.
Check that the reading on the respiratory pressure meter is within 3%
of the manometer reading.
If adjustment is required then the following procedure must be
followed.
The calibration may only be adjusted in a positive direction as the
meter monitors the peak pressure value. Therefore if the respiratory
pressure meter reading was greater than the manometer reading the
calibration screw must be turned anti-clockwise a few turns before
calibration is attempted.
Serial
port
Calibration
screw
Connect the respiratory pressure meter to the manometer as
previously shown. Fill the syringe to obtain the required negative
All manuals and user guides at all-guides.com
Summary of Contents for MicroRPM
Page 6: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 20: ...All manuals and user guides at all guides com...
Page 35: ...31 3 2 4 6 8 2 9 1 SNIP MIP MEP OFF 3 1 7 4 5 All manuals and user guides at all guides com...
Page 60: ...All manuals and user guides at all guides com...
Page 64: ...59 1 P max OFF MIP MEP All manuals and user guides at all guides com...
Page 65: ...60 0 TLC 2 OFF PI max OFF MIP MEP RV 2 All manuals and user guides at all guides com...
Page 66: ...61 4 OFF SNIP All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 67: ...62 0 SNIP FRC OFF BAT MicroRPM PUMA RS 232 All manuals and user guides at all guides com...
Page 68: ...63 MicroRPM 15 CareFusion 65 All manuals and user guides at all guides com...
Page 69: ...64 MicroRPM SNIP 200 300 MicroRPM 3 MicroRPM All manuals and user guides at all guides com...
Page 75: ...Notes All manuals and user guides at all guides com...