25
Cleaning
The respiratory pressure meter requires no routine maintenance or
servicing and is protected from contamination by the bacterial filter
supplied. However the mouthpiece and nasal probes may be
immersed in cold sterilising solutions such as PeraSafe for a period not
exceeding 15 minutes. Rinse with distilled water, drain, and allow to
dry before reassembly. A flow of air through the nasal probe
connecting tube may be introduced to shorten the drying time.
CareFusion can supply a pump for this purpose – see
Consumables/Supporting products, page 27. PeraSafe is available from
CareFusion in convenient 81g, plastic containers.
Calibration
Calibration is factory set and should remain stable indefinitely.
However, calibration may be checked by connecting the device to a
manometer as shown below:
Manometer
cm H O
2
0
2
1
3
6
4
5
8
7
9
Syringe
10
Connecting
line
Female
Luer
T Piece
Nasal plug
adapter
All manuals and user guides at all-guides.com
Summary of Contents for MicroRPM
Page 6: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 20: ...All manuals and user guides at all guides com...
Page 35: ...31 3 2 4 6 8 2 9 1 SNIP MIP MEP OFF 3 1 7 4 5 All manuals and user guides at all guides com...
Page 60: ...All manuals and user guides at all guides com...
Page 64: ...59 1 P max OFF MIP MEP All manuals and user guides at all guides com...
Page 65: ...60 0 TLC 2 OFF PI max OFF MIP MEP RV 2 All manuals and user guides at all guides com...
Page 66: ...61 4 OFF SNIP All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 67: ...62 0 SNIP FRC OFF BAT MicroRPM PUMA RS 232 All manuals and user guides at all guides com...
Page 68: ...63 MicroRPM 15 CareFusion 65 All manuals and user guides at all guides com...
Page 69: ...64 MicroRPM SNIP 200 300 MicroRPM 3 MicroRPM All manuals and user guides at all guides com...
Page 75: ...Notes All manuals and user guides at all guides com...