25
Cleaning
Disinfection of contaminated parts is only effective after having them
carefully pre-cleaned. CareFusion recommend the tested solution of
PeraSafe sterilizing powder (36-SSC5000A) for pre-cleaning and
disinfection. If a different solution is used, please follow the given
manufacturer’s instructions.
CAUTION:
Switch off the device and always unplug your BabyCO
meter from the computer before cleaning.
When using the disposable cardboard mouthpiece with one-way
valve under the prerequisite that the patient was instructed only to
exhale, the mouthpiece adapter and the CO meter surface have to
be cleaned.
The mouthpiece adapter may be cleaned and sterilised by immersion
in Perasafe (36-SSC5000A) or other cold sterilising solutions.
Exposed surfaces of the meter may be wiped with a clean, damp
cloth or a cloth that has been immersed in a sterilising solution. It is
recommended that this procedure is performed after each use and
all used cardboard mouthpieces discarded.
Important note:
The sensor surface must not be wiped with any
aqueous solutions and
must not
be exposed to solvents e.g. alcohol
or permanent damage may result
CAUTION:
Do not attempt to wash or immerse the BabyCO meter
in water or cleaning fluid, as there are electronic components inside
that will be permanently damaged.
Important note:
Cardboard mouthpieces must be disposed off
immediately after the use.
If there are changes on the material surfaces of either the unit or
mouthpiece adaptor (cracks, brittleness) the respective parts must
be disposed off.
Summary of Contents for BabyCO Meter
Page 2: ......
Page 10: ...8 ...
Page 41: ...39 ...
Page 42: ...40 ...
Page 47: ...45 6 5 4 1 2 3 CO FCOHb OFF ...
Page 50: ...48 Das BabyCO ist zur Verwendung durch Ärzte und medizinisches Fachpersonal vorgesehen ...
Page 79: ...77 ...
Page 80: ...78 ...
Page 114: ...112 ...
Page 146: ...144 ...
Page 181: ...179 ...
Page 182: ...180 ...
Page 216: ...214 ...
Page 221: ...219 6 5 4 1 2 3 CO FCOHb OFF ...
Page 254: ...252 ...
Page 264: ...262 将滑动开关推至 CO PPM 位置 等待设备显示清零 如下所示 将控制阀紧紧拧到气缸上 并连接气源 串行 定标 开关 ...